IEAP - Institute of Experimental and Applied Physics CTU - Czech Technical University in Prague
Česky English
IEAP - Institute of Experimental and Applied Physics CTU - Czech Technical University in Prague
CTU - Czech Technical University in Prague
Publication  > Articles in Impacted Journals  > '3D measurement of the radiation distribution in a water phantom in a hadron therapy beam'
3D measurement of the radiation distribution in a water phantom in a hadron therapy beam

Author
Solc Jaroslav IEAP

Year
2012

Scientific journal
JINST 7 C01085 doi:10.1088/1748-0221/7/01/C01085

Web


Abstract
Hadron therapy is a highly precise radio-therapeutic method with many advantages especially in cases when the tumour is close to sensitive organs where standard treatments cannot be used. For reliable treatment planning it is necessary to have calculation tools for maximization of the dose delivered to the targeted tissue and minimization of the dose outside of it. While the main physical processes in material irradiated by hadron beams are known, in reality the processes involved are complex so that analytical computations are impossible. Thus, the planning tools to incorporate simplified models and numerical approximations and an experimental method for high precision verification of the models within phantoms is desired. The development of sensitive, high resolution and online methods for measurement of the radiation environment inside of the irradiated object is the aim of this work. Such measurements are made possible by the resolving power of the state-of-the-art pixel detector Timepix. This quantum counting imaging device is able to record the characteristic shapes of the particle traces including their energies deposited in the detector. All these data recorded for each event allow to estimate the particle type, its energy and direction of flight. Event-by-event analysis is done using pattern recognition of the characteristic traces. The objective of the experiment is the detection and characterization of secondary radiation generated by the primary therapeutic beams in tissue equivalent material (water). Measurements were performed inside of a water phantom irradiated by a carbon beam at the Heidelberg Ion-Beam Therapy Center (HIT).

Cite article as:
J. Solc, "3D measurement of the radiation distribution in a water phantom in a hadron therapy beam", JINST 7 C01085 doi:10.1088/1748-0221/7/01/C01085 (2012)